VRS-100 SERIES HIGH RESISTANCE STANDARDS

Key Features

  • Set of nine, ten or eleven resistors – 1 kΩ, 10 kΩ, 100 kΩ,1 MΩ, 10 MΩ, 100 MΩ, 1 GΩ, 10 GΩ, 100 GΩ, 1 TΩ and 10 TΩ , open circuit (Depending on model)
  • High Accuracy – up to 20 ppm
  • Standard voltage: up to 5 kV (voltages of 10 kV are available)
  • Excellent for calibration of Meggers

Description

An economical set of precision high resistance standards designed for testing meggers or other instruments; 9 or 10 values up to 10 TΩ with special Kel-F isolation

  • Set of nine, ten or eleven resistors – 1 kΩ, 10 kΩ, 100 kΩ,1 MΩ, 10 MΩ, 100 MΩ, 1 GΩ, 10 GΩ, 100 GΩ, 1 TΩ and 10 TΩ , open circuit (Depending on model)
  • High Accuracy – as low as 20 ppm
  • Standard voltages: of 5 kVDC and optional 10 kVDC
  • Low Voltage Coefficient of Resistance even at 10 kVDC
  • Excellent TC – as low as 5 ppm/°C
  • Excellent stability – as low as 10 ppm/year
  • Excellent for Calibration of Hipot Testers and Megohmmeters

Two user versions available:

  • binding posts
  • rotary switch

Resistance values of 10 MΩ and less are calibrated with a Fluke 8508A Reference DMM. Resistance values above 10 MΩ are calibrated with a Keithley 6517B at 1000 Vdc.

In addition, all resistance values of 10 MΩ and greater are calibrated at 5 kVdc with a CMC* as good as 0.5%.

The standard VRS-100 features a maximum voltage of 5 kVDC for values above 10 M&Omega. An (-10kV) option is available for 10 kVDC maximum voltage for resistance values above 100 MΩ.

The VRS-100 Series can be custom made to put any set of resistance values from 0.01 Ω – 10 TΩ in 1 housing. For example we can put a 5 TΩ resistance standard in a unit. Units can be made with as few as 3 values and as many as 11 resistance values.

Please contact our sales staff if you have a custom requirement.

* See IET Labs scope of accreditation, A2LA Calibration Laboratory Certificate # 2073.01 for capabilities.

Related Products

PRS-300 PRECISION PROGRAMMABLE RESISTANCE DECADE

  • Resistance Range: 0.1 Ω to 20 MΩ
  • Precision Lab Grade Accuracy 70 ppm + 1 mΩ
  • Resolution: 1 μΩ ( 0.000 001 Ω) or 6 digits
  • To further improve accuracy to <20 ppm, for no additional cost, see the PRS-DMM
  • Programmable resistor with USB interface and Labview drivers
  • 5 year warranty

HPRS-HIPOT-LOAD (FLUKE 5320A-LOAD HIPOT REPLACEMENT) HIPOT CURRENT CALIBRATION LOAD RESISTORS

  • 8 Resistances from 10 kΩ to 5 MΩ standard
  • Standard accuracy: ±10% at 23°C
  • Power rating : up to 300 W at 100 kΩ
  • Voltage rating: 1,400 to 5,500 V depending upon resistance value
  • Continuous operation at rated power
  • Kel-F washers on all terminals for low leakage current and high dielectric strength
  • Custom values available for AR and QuadTech/Chroma hipot testers

ESI RS925D RESISTANCE DECADE

  • High accuracy – 20 ppm + 0.5 mΩ
  • High stability – 20 ppm + 0.5 mΩ
  • Resistance from 10 mΩ to 1.1 MΩ
  • 20 μΩ resolution
  • Precise fixed minimum resistance
  • Hermetically sealed, low inductance resistors
  • The first 4 decades can be individually trimmed for increase accuracy

ESI / TEGAM DB62 DECADE RESISTOR

  • Precision accuracy of ±(0.01% + 2 mΩ)
  • Resistance range from 0.01 Ω to 11.1111 MΩ
  • 4 models available, each covering six decades
  • High-performance solid silver-alloy contacts
  • Low zero resistance < 6mΩ

ESI SR1050 SERIES 1M & 10 M RESISTANCE TRANSFER STANDARD

  • Resistance Values 1 MΩ & 10 MΩ
  • Initial Accuracy… ± 25 or 30 ppm of nominal value matched within 10 ppm
  • Transfer Accuracy… ± 2 ppm (Typical Short-term Repeatability)
  • Calibration Accuracy… ± 10 or 15 ppm

ESI SR1010 RESISTANCE TRANSFER STANDARD

  • Resistance Transfer Standards available in the following values 1 Ω, 10 Ω, 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ
  • Initial Accuracy… ± 20PPM of nominal value matched within 10PPM
  • Transfer Accuracy… ± (1PPM + 0.1uΩ at parallel value) for 100:1
  • Transfer Accuracy… ± (1PPM + 1uΩ at series parallel value) for 10:1
  • Long Term Accuracy… ± 50PPM of nominal value
  • Calibration Accuracy… ± 10PPM

PRS-200 SERIES PROGRAMMABLE RESISTANCE DECADE BOX

  • IEEE, RS232 or Ethernet Programmable Decade Resistor
  • Wide range and resolution: 100 mΩ to 100 M&Omega with 1 mΩ resolution
  • Available at accuracies of : 0.01%, 0.02%, 0.05%, 0.1%, 0.5%, and 1%
  • Special RTD and custom configurations
  • Dual / Combo rackmount models ( examples: Resistance & Capacitance or Resistance & Inductance)

ESI SR1030 SERIES OIL IMMERSED TRANSFER STANDARD

  • Resistance Values of 1 Ω – 100 kΩ
  • Initial Accuracy… ± 20 PPM of nominal value matched within 10 PPM
  • Transfer Accuracy… ± (1 PPM + 0.1uΩ at parallel value) for 100:1
  • Long Term Accuracy… ± 50 PPM of nominal value
  • Calibration Accuracy… ± 10 PPM